Ordering Recommendation

New York DOH Approval Status

Testing is not New York state approved. Specimens from New York clients will be sent out to a New York state-approved laboratory.

Specimen Required

Patient Preparation
Collect

Serum separator tube (SST).

Specimen Preparation

Separate serum from cells ASAP or within 2 hours of collection. Transfer 0.5 mL serum to an ARUP standard transport tube. (Min: 0.25 mL).

Refer to "Allergen Specimen Collection Instructions" at www.aruplab.com/testing/resources/specimen.

Storage/Transport Temperature

Refrigerated.

Unacceptable Conditions

Hemolyzed, icteric, or lipemic specimens.

Remarks
Stability

After separation from cells: Ambient: 48 hours; Refrigerated: 2 weeks; Frozen: 1 year

Methodology

Quantitative ImmunoCAP Fluorescent Enzyme Immunoassay

Performed

Sun

Reported

1-8 days

Reference Interval

Less than 9.77 mcg/mL

Interpretive Data

Values less than 2.00 mcg/mL represent absent or undetectable levels of allergen-specific IgG antibody.

Compliance Category

Laboratory Developed Test (LDT)

Note

The units of measure mcg/mL and mgA/L are interchangeable. 1 mg/L = 1000 mcg/1000 mL

Hotline History

N/A

CPT Codes

86001

Components

Component Test Code* Component Chart Name LOINC
0097316 Allergen, Fungi/Mold, M. racemosus IgG 35549-5
* Component test codes cannot be used to order tests. The information provided here is not sufficient for interface builds; for a complete test mix, please click the sidebar link to access the Interface Map.

Aliases

Allergen, Fungi and Molds, Mucor racemosus IgG