Ordering Recommendation

Acceptable test to detect and subtype protein C deficiency. Do not order if individual has been on warfarin therapy in the previous 2-4 weeks.

New York DOH Approval Status

This test is New York state approved.

Specimen Required

Patient Preparation
Collect

Lt. blue (sodium citrate). Refer to Specimen Handling at aruplab.com for hemostasis/thrombosis specimen handling guidelines.

Specimen Preparation

Transfer 2 mL platelet-poor plasma to an ARUP standard transport tube. (Min: 1 mL)

Storage/Transport Temperature

CRITICAL FROZEN. Separate specimens must be submitted when multiple tests are ordered.

Unacceptable Conditions

Serum. EDTA plasma, clotted or hemolyzed specimens.

Remarks
Stability

Ambient: 4 hours; Refrigerated: Unacceptable; Frozen at -20°C: 3 months, at -70°C: 6 months

Methodology

Electromagnetic Mechanical Clot Detection/Enzyme-Linked Immunosorbent Assay

Performed

Sun-Sat

Reported

1-5 days

Reference Interval

By report

Interpretive Data

Refer to report.

Compliance Category

FDA

Note

If protein C functional is decreased, then Protein C, Total Antigen, will be added. Additional charges apply.

Hotline History

N/A

CPT Codes

85303; if reflexed, add 85302

Components

Component Test Code* Component Chart Name LOINC
0030113 Protein C Functional 27819-2
* Component test codes cannot be used to order tests. The information provided here is not sufficient for interface builds; for a complete test mix, please click the sidebar link to access the Interface Map.

Aliases

  • Protein C Profile
  • Protein C, functional and total
Protein C, Functional with Reflex to Protein C, Total Antigen